Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: HP 974A: (6 month) CAL VER /5500 DATE: 08-Oct-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 57 NUMBER OF LINES: 190 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P J F W 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 23 +/-5 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Turn the unit ON. 1.007 DISP Connect the 5500A to the UUT as follows: 1.007 DISP [32] 5500A NORMAL HI to UUT VOLT-OHM-DIODE 1.007 DISP [32] 5500A NORMAL LO to UUT COM 1.008 HEAD {AC VOLTAGE} 1.009 HEAD AC VOLTAGE: {500mV Range} 1.010 5500 500 480.00mV 1% 0.3U 20H SI 2W 2.001 5500 500 480.00mV 0.7% 0.3U 1kH SI 2W 3.001 5500 500 480.00mV 2% 0.5U 20kH SI 2W 4.001 HEAD AC VOLTAGE: {5V Range} 4.002 5500 5 4.8000V 1% 0.003U 20H SI 2W 5.001 5500 5 4.8000V 0.5% 0.003U 1kH SI 2W 6.001 5500 5 4.8000V 1% 0.004U 20kH SI 2W 7.001 5500 5 4.8000V 2% 0.007U 40kH SI 2W 8.001 5500 5 4.8000V 3% 0.03U 50kH SI 2W 9.001 HEAD AC VOLTAGE: {50V Range} # 48V 20Hz test reduced to 30V 20Hz due to 5500A range limit. 9.002 5500 50 30.000V 1% 0.03U 20H SI 2W 10.001 5500 50 48.000V 0.5% 0.03U 1kH SI 2W 11.001 5500 50 48.000V 1% 0.04U 20kH SI 2W 12.001 5500 50 19.000V 2% 0.07U 40kH SI 2W 13.001 5500 50 19.000V 3% 0.3U 50kH SI 2W 14.001 HEAD AC VOLTAGE: {500V Range} 14.002 5500 500 480.00V 1% 0.3U 50H SI 2W 15.001 5500 500 480.00V 0.5% 0.3U 1kH SI 2W # 480V 20kHz test reduced to 300V 20kHz due to 5500A range limit. 16.001 5500 500 300.00V 1% 0.4U 20kH SI 2W 17.001 HEAD {FREQUENCY} 17.002 DISP Press the button twice to select dBm. 22.004 5500 -23.00D 0.2U 5kH SI 2W 23.001 5500 -23.00D 0.5U 15kH SI 2W 24.001 5500 -3.00D 0.15U 5kH SI 2W 25.001 5500 -3.00D 0.3U 15kH SI 2W 26.001 5500 50.00D 0.1U 5kH SI 2W 27.001 5500 50.00D 0.2U 15kH SI 2W 28.001 5500 30.00D 0.5U 40kH SI 2W 29.001 5500 30.00D 1U 40kH SI 2W 30.001 5500 59.00D 0.15U 1kH SI 2W 31.001 HEAD {DC VOLTAGE} 31.002 HEAD DC VOLTAGE: {5V Range} 31.003 5500 5 4.8000V 0.05% 0.0002U 2W 32.001 HEAD DC VOLTAGE: {50V Range} 32.002 5500 50 48.000V 0.05% 0.002U 2W 33.001 HEAD DC VOLTAGE: {500V Range} 33.002 5500 500 480.00V 0.05% 0.02U 2W 34.001 HEAD DC VOLTAGE: {1000V Range} 34.002 5500 1000 1000.0V 0.05% 0.2U 2W 35.001 HEAD {DC+AC VOLTAGE} 35.002 RNG 5V 35.003 DISP Press the button once to select Continuity. 48.003 5500 50.0Z S 2W 48.004 EVAL Does {the BUZZER SOUND in Continuity Mode}? 49.001 HEAD {LOW DC CURRENT} 49.002 RNG 500uA 49.003 DISP Open the Current Shutter door to expose the 49.003 DISP current input connectors. 49.004 DISP Connect 5500A to the UUT as follows: 49.004 DISP [32] 5500A AUX HI to UUT mA-uA 49.004 DISP [32] 5500A AUX LO to UUT COM 49.005 HEAD LOW DC CURRENT: {500uA Range} 49.006 5500 500 480.000uA 0.3% 0.02U 2W 50.001 HEAD LOW DC CURRENT: {50mA Range} 50.002 5500 50 48.0000mA 0.3% 0.002U 2W 51.001 HEAD LOW DC CURRENT: {500mA Range} 51.002 5500 500 480.000mA 0.3% 0.02U 2W 52.001 HEAD {LOW AC CURRENT} 52.002 DISP Press the button to select DC Current. 56.002 HEAD {HIGH DC CURRENT} 56.003 HEAD HIGH DC CURRENT: {10A DC Range} 56.004 ASK- R 56.005 5500 10 9.0000A 0.7% 0.002U 2W 57.001 END